(Ⅰ)X-ray Penetrability and Image Quality
5KW C-arm
Limited by maximum output current and upper voltage limits, suitable for thin layers and low-density tissues;
Standard body types (BMI < 26) can achieve normal imaging of limbs, hands/feet, shoulders/elbows, and simple lumbar vertebrae;
Limitations: For obese patients (BMI ≥ 28), thick femurs, full spine anteroposterior/lateral views, pelvis, and thick soft tissue areas, images appear grayish, with heavy graininess, and screws and pedicle boundaries are blurred;
When imaging multiple metal implants such as plates and long intramedullary nails, severe metal artifacts occur, making it easy to obscure screw alignment.
15KW C-arm
Sufficient instantaneous output allows simultaneous increase in kV and mA, significantly enhancing penetration through thick tissues;
Obese individuals, pelvis, full spine, thigh femur, and thoracoabdominal interventions can obtain low-noise, high-definition images, with clear visibility of trabecular bone, screw threads, and minor fracture lines;
Imaging of metal implants has stronger layering, allowing distinction between bone and internal fixation boundaries even with overlapping plates, reducing the need for repeated exposures and retakes.
(II) Continuous Surgical Tolerance (X-ray Tube Heat Dissipation)
5KW Model
The anode heat capacity is relatively small, and the heat dissipation efficiency is average:
· Suitable for short-duration surgeries: simple fracture reduction, small needle knife procedures, injections, hand and foot minimally invasive surgeries, and outpatient wound debridement and foreign body removal;
· During prolonged continuous fluoroscopy (>10 minutes) or continuous large-volume spot radiography, the X-ray tube overheating protection is frequently triggered. The equipment automatically reduces power, causing image darkening, and may even shut down for cooling, severely disrupting the surgical rhythm;
· Not suitable for a large number of consecutive complex orthopedic surgeries.
15KW Model
Equipped with a standard high-capacity oil-cooled X-ray tube, with significantly higher anode heat capacity:
· Supports long-duration continuous fluoroscopy: spinal screw fixation, hip joint replacement, complex pelvic fractures, interventional ablation, and multiple consecutive surgeries;
· During surgery, repeated spot radiography and multi-angle oblique fluoroscopy can be performed without the X-ray tube overheating. There is no problem of mid-surgery shutdown for cooling, greatly improving surgical smoothness;
· Preferred for tertiary hospital operating rooms and trauma centers with more than 20 daily surgeries.
(III) Exposure Frequency and Radiation Dose to Patients and Medical Staff
Many people mistakenly believe that higher power radiation is greater, but the actual clinical logic is the opposite:
1.5KW Limitations: Insufficient penetration. To see the image clearly, fluoroscopy time must be extended and repeated exposures performed, resulting in more total exposures and higher cumulative radiation doses for both patients and medical staff; for obese patients, images are blurred, leading doctors to repeatedly step on the foot pedal for fluoroscopy positioning, doubling the radiation.
2
15KW Advantages: A single exposure yields a qualified image without the need for repeated fluoroscopy. Total exposure duration and number of shots are fewer. With the aid of automatic dose adjustment algorithms, skin dose per exposure is controllable under equivalent image quality, resulting in lower overall radiation and compliance with radiological protection standards.